C9AD026 L Start, Current Program Config Report EBAE006 L 1,P1 06AF02B L 2,P1\X\Test Type=Diff Press Decay-dP c 91B0034 L 3,P1\X\= a 4CB1034 L 4,P1\X\=Measures the Pressure Loss (dP), a E1B2034 L 5,P1\X\=determined from the differential loss a 79B3034 L 6,P1\X\=between the test part and the a 49B4034 L 7,P1\X\=reference volume over the duration of a C6B5034 L 8,P1\X\=the test timer. The result is a 76B6034 L 9,P1\X\=presented in units of delta pressure. a 29B7021 L 10,P1\R\Tool Ext 1=2.50 sec f F8B801D L 11,P1\R\Charge=5.00 sec f D6B901B L 12,P1\R\Fill=5.00 sec f E8BA021 L 13,P1\R\Stabilize=10.00 sec f 45BB01F L 14,P1\R\Isolation=50.00 % f 6CBC01C L 15,P1\R\Test=10.00 sec f E3BD01E L 16,P1\R\Exhaust=0.50 sec f DDBE021 L 17,P1\R\Tool Ret 1=1.00 sec f 82BF029 L 18,P1\P\Charge Min Press=0.400 mbar f 3EC002E L 19,P1\P\Charge Target Press=140.000 mbar f 77C102C L 20,P1\P\Charge Max Press=6894.757 mbar f 49C202A L 21,P1\P\Minimum Pressure=95.000 mbar f 27C302A L 22,P1\P\Target Pressure=100.000 mbar f 39C402B L 23,P1\P\Maximum Pressure=120.000 mbar f E5C502A L 24,P1\P\Target Press Window=100.00 % f DBC602A L 25,P1\S\Low Limit Loss=-6894.758 dPa f 3EC7028 L 26,P1\S\High Limit Loss=90.000 dPa f D5C8023 L 27,P1\S\Decay Direction=Loss c F8C9024 L 28,P1\S\Test Evaluation=F/P/F c 74CA023 L 29,P1\S\Loss Offset=0.000 dPa f 01CB01D L 30,P1\S\EDC Enabled=No c 72CC021 L 31,P1\S\EDC Percentage=25% c 26CD01E L 32,P1\S\EDC Quantity=30 i 4BCE01F L 33,P1\M\Program Name=H R a CACF023 L 34,P1\M\Barcode Required=Yes c A9D0024 L 35,P1\M\Barcode Method=Unique c 20D1024 L 36,P1\M\Copy to Target Prog=1 i 47D201E L 37,P1\M\Copy Program=No c 4AD3023 L 38,P1\M\Retain Part Press=No c 11D4026 L 39,P1\M\Malfunction Eval=Reject c A1D5030 L 40,P1\M\Pressure Select=Pressure Source 1 c 79D6023 L 41,P1\M\Leak Std Select=None c 8BD7022 L 42,P1\M\ERC Method=Disabled c 63D802A L 43,P1\M\Prefill Method=Fixed Charge c 52D902A L 44,P1\M\Pressure Correction=Enabled c B0DA022 L 45,P1\M\Exhaust Method=Dump c 49DB024 L 46,P1\A\Autorun Relax=3.00 sec f 18DC026 L 47,P1\A\Autorun Method=Standard c 57DD025 L 48,P1\A\Autorun Cycle Count=10 i 99DE020 L 49,P1\A\Autorun Enable=No c 58DF01C L 50,P1\U\Time Unit=sec c 9BE0022 L 51,P1\U\Time Precision=X.XX c F9E1021 L 52,P1\U\Pressure Unit=mbar c F4E2027 L 53,P1\U\Pressure Precision=X.XXX c FFE3022 L 54,P1\U\d Pressure Unit=dPa c F2E4026 L 55,P1\U\d Press Precision=X.XXX c 24E501C L 56,P1\I\I/O ID=786561 a 99E601E L 57,P1\I\1=Start Channel c 49E701D L 58,P1\I\2=Part Present c 1EE801B L 59,P1\I\3=Unassigned c 66E901B L 60,P1\I\4=Unassigned c 48EA01B L 61,P1\I\5=Unassigned c BEEB01B L 62,P1\I\6=Unassigned c ECEC01B L 63,P1\I\7=Unassigned c BFED01C L 64,P1\I\8=Constrained c A5EE01B L 65,P1\I\9=Unassigned c 2EEF01C L 66,P1\I\10=Unassigned c 46F001C L 67,P1\I\11=Unassigned c DDF101C L 68,P1\I\12=Unassigned c FBF201C L 69,P1\O\I/O ID=786561 a D5F301D L 70,P1\O\1=Press Select c 27F401F L 71,P1\O\2=Program Reject c 71F501C L 72,P1\O\3=Test Passed c 1BF601E L 73,P1\O\4=Tool Extend 1 c FAF701B L 74,P1\O\5=Unassigned c 52F801B L 75,P1\O\6=Unassigned c 08F901C L 76,P1\O\7=Constrained c 5EFA01F L 77,P1\O\8=Program Reject c D6FB01C L 78,P1\O\9=Constrained c 13FC01D L 79,P1\O\10=Constrained c 70FD01C L 80,P1\O\11=Unassigned c 98FE01C L 81,P1\O\12=Unassigned c 11FF020 L 82,P1\T\Tooling Option=On c EF00022 L 83,P1\T\Number of Motions=1 c 6C01025 L 84,P1\T\Two Inputs to Start=No c DE0202C L 85,P1\T\Part Present Check=Test Start c E60301E L 86,P1\T\Motion Number=1 c 5A04024 L 87,P1\T\Retract on Reject=Yes c 690501D L 88,P1\T\Part Mark=None c C306026 L 89,P1\T\Tooling Feedback 1=None c 8507025 L Stop, Current Program Config Report 2E08028 V I\S\RS232 1 REPORT=Cur Prog Config c