TRANSDUCER - array of TRANSDUCERS - Instrument Pressure \d # As Found - Master Value \d # Gauge Reading Set Point - Note: IP1 = MV1 Once selected, prompt for "As Found", "As Left", or "Both" - If as left, proppt for as found file. etc etc. - If both, continue on with logic. If Transducer Name contains "CUSTOM" - prompt for the name in an input box (change to Type) Part Number: "CTS A12-221" Value "250CCM" Data to always enter in - Instrument - Customer Name - Customer Address - Control Number - Serial Number - Accuracy (once, but applies to the full List) - Barometric Pressure - Temperature - Humidity Auto Data - Date [Verify Date] - Part Number "middle number from Part Name/Type variable" Calculations: etc - High Limit/ Low Limit eg: (SPAN*1.0005 + column) (SPAN*0.9995 - column) - Delta = abs(gauge reading) - abs(asleft|asfound) - PSIA - PRESSURE - SCCM - FLOW - TRANSDUCER1 Will ALWAYS be pressure CONFIG: - Mensor Numbers etc - Cal device, cal serial, Cal cal date, cal due date # HARDWARE CALIBRATION - "Mass Flow Trans" look at `Instrument Flow` and `Master Reading` - "Pressure Transducer" `Instrument Pressure` and `Master Value` - Always take the SECOND `Master Value`