import {ONE_NEW_LINE, TWO_NEW_LINES} from "./utils/constants.js";
const isInRange = (value, masterValue) => {
return (masterValue["Low Limit"] <= value && value <= masterValue["High Limit"]);
}
const calculateDelta = (value, lowLimit) => {
return Math.abs(lowLimit - value);
}
function outOfTolerance(readings) {
// Calculate Out of Tolerances
for (const reading of readings) {
@ -148,6 +147,11 @@ const parseCalibrationData = (text, accuracy) => {
}
function parseHardwareCalibration(content, accuracy) {
// hack because we can't be sure that the file will end in two newlines, so might as well force it to add two
// this way if there's only zero or one, we can still regex match on that
content += "\r\n\r\n"
// Replace the newlines consistently on windows/linux
content = content.replace(/\r\n/g, "\r").replace(/\n/g, "\r")
const [instrument, ports] = content.split("|| Hardware Calibration Report ||");
const instrumentInfo = parseInstrumentInfo(instrument);
@ -157,5 +161,6 @@ function parseHardwareCalibration(content, accuracy) {
}
export default function ParseHardwareCalibration(content, accuracy) {
return parseHardwareCalibration(content, accuracy);
}

View File

@ -0,0 +1,134 @@
|| Instrument Info ||
Instrument Name Chassis2 Adult
Serial Number BBP1736
Time 11:31:39
Date 07/12/2022
|| Hardware Calibration Report ||
Hardware 0
===============================================================
Test Port 1
===========================================================
Mass Flow Trans
===========================================================
Transducer Custom
Input AIN 2
Setpoint Pressure 1 25.000 psig
Analog Value 1 8688254 cnts
Master Value 1 0.041 sccm
Linearity 0.129 %
Last Cal Date 12/07/2020
Last Cal Time 10:35:20
Xdcr Filter Code Median
Xdcr Typ Residual 108.000
Xdcr Zero LL 0.000 cnts
Xdcr Zero HL 0.000 cnts
Xdcr Zero Lwin 0.000
Xdcr Zero Hwin 0.000
Xdcr Analog Min 0.000 cnts
Xdcr Analog Max 0.000 cnts
Xdcr Base Min 0.000
Xdcr Base Max 0.000
Setpoint Pressure 1 5.800 psig
Instrument Pressure 1 5.824 psig
Instrument Flow 1 0.038 sccm
Master Value 1 0.000 sccm
Master Reading 1 0.000 sccm
Verify Date 07/12/2022
Verify Time 10:54:56
Pressure Transducer
===========================================================
Transducer CTS D34-441 45PSIA
Input AIN 1
Setpoint Pressure 1 0.000 psig
Analog Value 1 10087020 cnts
Master Value 1 0.001 psig
XDCR Analog ATM 10086749
Atm Pressure -0.000 bar
Linearity 0.082 %
Last Cal Date 12/07/2020
Last Cal Time 09:03:38
Xdcr Filter Code Med, J-Filter
Xdcr Typ Residual 80.000
Xdcr Zero LL 9654557.000 cnts
Xdcr Zero HL 10485333.000 cnts
Xdcr Zero Lwin -2.501 psig
Xdcr Zero Hwin 2.499 psig
Xdcr Analog Min 5699952.000 cnts
Xdcr Analog Max 15110248.000 cnts
Xdcr Base Min -4294967.000
Xdcr Base Max 8589935.000
Setpoint Pressure 1 0.000 psig
Instrument Pressure 1 0.000 psig
Master Value 1 0.000 psig
Verify Date 07/12/2022
Verify Time 08:14:59
Test Port 2
===========================================================
Mass Flow Trans
===========================================================
Transducer Custom
Input AIN 4
Setpoint Pressure 1 25.000 psig
Analog Value 1 8689552 cnts
Master Value 1 0.054 sccm
Linearity 0.166 %
Last Cal Date 12/07/2020
Last Cal Time 10:50:52
Xdcr Filter Code Median
Xdcr Typ Residual 83.000
Xdcr Zero LL 0.000 cnts
Xdcr Zero HL 0.000 cnts
Xdcr Zero Lwin 0.000
Xdcr Zero Hwin 0.000
Xdcr Analog Min 0.000 cnts
Xdcr Analog Max 0.000 cnts
Xdcr Base Min 0.000
Xdcr Base Max 0.000
Setpoint Pressure 1 5.800 psig
Instrument Pressure 1 5.780 psig
Instrument Flow 1 0.049 sccm
Master Value 1 0.000 sccm
Master Reading 1 0.000 sccm
Verify Date 07/12/2022
Verify Time 11:01:54
Pressure Transducer
===========================================================
Transducer CTS D34-441 45PSIA
Input AIN 3
Setpoint Pressure 1 0.000 psig
Analog Value 1 10050243 cnts
Master Value 1 0.001 psig
XDCR Analog ATM 10050055
Atm Pressure -0.000 bar
Linearity 0.082 %
Last Cal Date 12/07/2020
Last Cal Time 09:11:34
Xdcr Filter Code Med, J-Filter
Xdcr Typ Residual 59.000
Xdcr Zero LL 9654557.000 cnts
Xdcr Zero HL 10485333.000 cnts
Xdcr Zero Lwin -2.501 psig
Xdcr Zero Hwin 2.499 psig
Xdcr Analog Min 5699952.000 cnts
Xdcr Analog Max 15110248.000 cnts
Xdcr Base Min -4294967.000
Xdcr Base Max 8589935.000
Setpoint Pressure 1 0.000 psig
Instrument Pressure 1 -0.000 psig
Master Value 1 0.000 psig
Verify Date 07/12/2022
Verify Time 08:21:53

View File

@ -1,7 +1,8 @@
const fs = require('fs');
import {readFiles} from "../utils/file_utils.js";
import ParseHardwareCalibration from "../Hardware"
// const file = fs.readFileSync("src/parsers/__tests__/hardware_calibration/hardware_calibration.txt", 'utf8')
describe("Test for all files", () => {
@ -13,3 +14,74 @@ describe("Test for all files", () => {
})
}
});
describe("Testing proper shape", () => {
test("It outputs an array of arrays", () => {
const content = fs.readFileSync("src/parsers/__tests__/hardware_calibration/smallest.txt", 'utf8');
const calibrations = ParseHardwareCalibration(content, 0.05);
expect(calibrations).toEqual(
{
"Calibration": {
"Test Port 1": {
"Mass Flow Trans": {
"Gauge Reading": [{
"Delta": 0,
"High Limit": 0,
"In Range": false,
"Low Limit": 0,
"Master Value": 0,
"Out Of Tolerance": 0,
"Unit": "sccm",
"Value": 0
}], "Name": "Custom", "Verify Date": "07/12/2022", "Verify Time": "10:54:56"
},
"Pressure Transducer": {
"Gauge Reading": [{
"Delta": 2250,
"High Limit": 2250,
"In Range": false,
"Low Limit": -2250,
"Master Value": 0,
"Out Of Tolerance": 2250,
"Unit": "psig",
"Value": 0
}], "Name": "45PSIA", "Verify Date": "07/12/2022", "Verify Time": "08:14:59"
}
},
"Test Port 2": {
"Mass Flow Trans": {
"Gauge Reading": [{
"Delta": 0,
"High Limit": 0,
"In Range": false,
"Low Limit": 0,
"Master Value": 0,
"Out Of Tolerance": 0,
"Unit": "sccm",
"Value": 0
}], "Name": "Custom", "Verify Date": "07/12/2022", "Verify Time": "11:01:54"
},
"Pressure Transducer": {
"Gauge Reading": [{
"Delta": 2250,
"High Limit": 2250,
"In Range": false,
"Low Limit": -2250,
"Master Value": 0,
"Out Of Tolerance": 2250,
"Unit": "psig",
"Value": -0
}], "Name": "45PSIA", "Verify Date": "07/12/2022", "Verify Time": "08:21:53"
}
},
},
"Instrument": {
"Date": "07/12/2022",
"Instrument Name": "Chassis2 Adult",
"Serial Number": "BBP1736",
"Time": "11:31:39"
}
}
);
})
});

View File

@ -15,7 +15,7 @@ describe("Test for all files", () => {
expect(transducer).toHaveProperty("Gauge Reading")
expect(transducer).not.toHaveProperty("Master Value")
expect(transducer["Gauge Reading"].length).toBeGreaterThan(1);
expect(transducer["Gauge Reading"].length).toBeGreaterThan(0);
}
});
}
@ -50,6 +50,64 @@ describe("Testing Errors", () => {
ParseTransducer(`|| Transducer Verify Report ||\nTRANSDUCER1\n===============================================================\nTransducer 1 CTS D34-442 115FigNewtons`, 0);
}
expect(e).toThrowError(Error("Unknown Type of Test, do not know unit: FIGNEWTONS"))
})
})
describe("Testing proper shape", () => {
test("It outputs an array of arrays", () => {
const content = fs.readFileSync("src/parsers/__tests__/transducer_verify/smallest.txt.txt", 'utf8');
const transducers = ParseTransducer(content, 0.05);
expect(transducers).toEqual(
[
{
Instrument: {
Accuracy: 0.0005,
Value: 115,
Unit: 'PSIA',
'Part Number': 'CTS D34-442',
'Limit ABS': 57.5,
'Transducer Name': 'Transducer 1',
'Transducer Type': 'Pressure',
'Verify Date': '07/20/22',
'Verify Time': '11:20:26'
},
'Gauge Reading': [
{
"Delta": 57.5,
"High Limit": 57.5,
"In Range": true,
"Low Limit": -57.5,
"Master Value": 0,
"Out Of Tolerance": 0,
"Value": 0,
},
]
},
{
Instrument: {
Accuracy: 0.0005,
Value: 250,
Unit: 'SCCM',
'Part Number': 'CTS A12-221',
'Limit ABS': 125,
'Transducer Name': 'Transducer 2',
'Transducer Type': 'Flow',
'Verify Date': '07/15/21',
'Verify Time': '14:55:10'
},
'Gauge Reading': [
{
"Delta": 125,
"High Limit": 125,
"In Range": true,
"Low Limit": -125,
"Master Value": 0,
"Out Of Tolerance": 0,
"Value": -0,
},
]
}
]
)
})
})
});

View File

@ -0,0 +1,21 @@
|| Transducer Verify Report ||
TRANSDUCER1
===============================================================
Transducer 1 CTS D34-442 115PSIA
Setpoint Pressure 1 0.000 psig
Instrument Pressure 1 0.000 psig
Master Value 1 0.000 psig
Verify Date 07/20/22
Verify Time 11:20:26
TRANSDUCER2
===============================================================
Transducer 2 CTS A12-221 250SCCM
Setpoint Pressure 1 20.000 psig
Instrument Pressure 1 20.154 psig
Instrument Flow 1 -0.082 sccm
Master Reading 1 0.000 sccm
Master Value 1 0.000 sccm
Verify Date 07/15/21
Verify Time 14:55:10